Calibration – GF Signet 2610 Process Optical Dissolved Oxygen Sensor User Manual

Page 2

Advertising
background image

2

2610 DO Sensor

Introduction

Unpacking the Sensor

1. Remove the RDO

®

sensor from the box and other

packaging materials.

2. Unscrew the nose cone from the sensor and

remove the red protective dust cap from the
sensor. Save the dust cap for later use.

3. Remove the sensor cap from its shipping/storage

sleeve.

4. Align the two lines on the cap with the fl at part of

the sensor and fi rmly press (DO NOT TWIST) the
cap onto the sensor until it seals over the probe
body.

• CAUTION:

Twisting the sensor cap can

permanently damage both the cap and the
sensor.

• Avoid allowing moisture, including atmospheric

humidity, inside the cap. Keep the cap in its
sealed packaging until you are ready to install
it. Install promptly. Make sure that O-ring
grooves are dry and the O-ring is not rolled or
pinched inside the cap.

• The cap's lifetime is 1 year after the fi rst

reading has been taken. Install by the date
printed on the packaging.

5. Reattach the nose cone

Dust cap

Nose cone

Shipping/Storage sleeve

Calibration

Sensor

Align flat side of the tip
with lines on cap

Calibration is not required. The unit, as shipped from
the factory, will measure within 2% of reading for the
life of the sensor cap.

Replacing the sensor cap will keep the reading within
2% accuracy.