GF Signet 2610 Process Optical Dissolved Oxygen Sensor User Manual

Page 8

Advertising
background image

8

2610 DO Sensor

1. Wire the 9900 and 3-2610-41 as shown in the diagram.

NOTE:

The wiring of the 3-2610-41 is nonstandard:

• 3-2610-41 Red wire is connected to 12 to 24 VDC
• 3-2610-41 White wire is connected to S

3

L Data

• 3-2610-41 Black wire is connected to VDC Ground
• A jumper wire must be connected between the VDC

Ground and S

3

L Gnd.

2. On the 9900 press and hold the ENTER key for 2

seconds.

3. Press

the

▼ key to select the INPUT menu item.

4. Press the Enter key to access the INPUT menus.
5. Press

the

▲ key to select the TYPE menu item.

6. Press

the

► key to change the

TYPE selection.

7. Enter the code.
8. Repeatedly press the ▼ key until the “4-20 mA INPUT”

menu item is fl ashing and then press ENTER.

9. The 9900 will prompt “All settings will be reset”. Press the ▲ key to select “Yes” and then press ENTER.
10. The 9900 will return to the View Mode display.
11. Press and hold the ENTER key for 2 seconds.
12. Press the ▼ key to select the INPUT menu item.
13. In the NAME menu item, press the ► key to change the displayed name from “4-20 mA INPUT” to a more

descriptive name (e.g., "DISSOLVED O2") and press ENTER when done.

14. Press the ▼ key to select SENSOR UNIT menu item.
15. Press the ► to change the label from UNIT to MG/L or PPM and press ENTER.
16. Press the ▼ key and ensure the 4 mA VALUE is set to 0.0000.
17. Press the ▼ key and change the 20 mA VALUE from 5.0000 to 20.000 and press ENTER.
18. Press both the ▲ and ▼ keys simultaneously to return to the Menu.
19. Press the ▼ to select the LOOP menu and press ENTER.
20. Set the 4 MA SET POINT to your desired value. The 2610 is factory set for a 0 to 20 mg/L output. Press

ENTER when done.

21. Press the ▼ key to select the 20 MA SET POINT and set to your desired value. The 2610 is factory set for a 0

to 20 mg/L output. Press ENTER when done.

22. Press both the ▲ and ▼ keys simultaneously to return to the Menu.
23. Press the ▼ key twice to select the OPTION menu and press ENTER.
24. Press the ▼ key twice to select the SET BAR MIN option. Change this option if desired. The 2610 is factory

set for a 0 to 20 mg/L output. Press ENTER when done.

25. Press the ▼ key to select the SET BAR MAX option. Change this option if desired. The 2610 is factory set for

a 0 to 20 mg/L output. Press ENTER when done.

26. Press both the ▲ and ▼ keys simultaneously to return to the Menu.
27. Enter the other menus and set up the unit as desired for your application.
28. Press both the ▲ and ▼ keys simultaneously to return to the View Menu.

Wiring and Set Up

3-2610-41 to 9900 Set Up

The 3-2610-41 Optical Dissolved Oxygen Sensor with S

3

L was designed to emulate a 4 to 20 mA current input

device, 8058, on the 9900 controller. This allows the 3-2610-41 to be backward compatible with all existing 9900
controllers.

RED

RED

BLACK

BLACK

WHITE

SHIELD

PWR–

D

ATA

GND

SHLD

V+

9900 S

3

L Inputs

9900 Power Inputs

LOOP–

LOOP+

PWR+

RED

Blue

Green

Brown

3-2610-41

No connection

V+

V-

24 VDC

Regulated

Power Supply

9900 Generation III supports 3-2610-41 direct connection.
Please refer to the 9900 product manual for setup instructions.