GF Signet 2610 Process Optical Dissolved Oxygen Sensor User Manual

Page 6

Advertising
background image

6

2610 DO Sensor

1. Connect the 3-2610-41, and 8900 as shown in the diagram

NOTE:

The wiring of the 3-2610-41 is non-standard:

• Red wire is connected to 12 to 24 VDC
• White wire is connected to S

3

L Data

• Black wire is connected to VDC Ground
• A jumper wire must be connected between the VDC Ground and S

3

L Gnd.

2. Press and hold Enter key for 5 seconds. The display will read “System Setup”.
3. Press Enter key.
4. Scroll to the channel that will be dissolved oxygen.
5. Press

the

► key.

6. Enter the password.
7. Repeatedly press the ▼ key until the lower

line of the display reads “Other (4-20)”.

8. Press Enter key.
9. The 8900 will warn that “Channel Data will

be Reset. Are you sure?” press the ▼ key so
that “Yes” is fl ashing and press Enter key.

10. Simultaneously press the ▲ and ▼ keys to

return to the Menu Directory.

11. Press the ▼ key to select the menu “Channel

Settings” and press Enter key to select.

12. Use the ▼ key to select the channel used in

step 4.

13. Press the ► key to change the label and

press Enter when done.

14. Press the ▼ key to select the abbreviation.
15. Press the ► key to change the label and

press Enter when done.

16. Press the ▼ key to select the Units.
17. Press the ► key to change the Units. The

2610 shipped with default units of mg/L, or
ppm, either can be used. Press Enter when
completed.

18. Press the ▼ key to select the 4mA Set Point.
19. Press the ► key to change the set point. By default the 4mA set point for the 2610 is 0.0. Press the Enter key

when complete.

20. Press the ▼ key to select the 20mA Set Point.
21. Press the ► key to change the set point. By default the 20 mA set point for the 2610 is 20.0. Press the Enter

key when complete.

22. Press the ▼ key to select the Decimal location..
23. Press the ► key and change the decimal position if desired. Press Enter when complete.
24. Simultaneously press the ▲ and ▼ keys to exit out of the channel menu.
25. Continue programming other options in the 8900 or simultaneously press the ▲ and ▼ keys to return to

View Mode.

Wiring and Set Up

3-2610-41 to 8900 Set Up

The 3-2610-41 Optical Dissolved Oxygen Sensor with S

3

L was designed to emulate a 4 to 20 mA current

input device, e.g., Signet 3-8058, on the 8900 controller. This allows the 3-2610-41 to be backward
compatible with all existing 8900 controllers.

I/O Module 3-8900.401-X

1

2

3

4

5

6

7

8

9

10

11

12

13

14

+5VDC (Black)

Freq. Input (Red)

GND (Shield)

+5VDC (Black)

Freq. Input 2 (Red)

S L (Red)

GND (White/Shield)

+5VDC (Black)

S L (Red)

GND (White/Shield)

3

3

Analog Output 1

Analog Output 2

(if applicable)

(if applicable)

Frequency

Input

1

Frequency

Input 2

OR

S3L

Input

2

S3L

Input

1

+

-

+

-

OUT

2 OUT

1

POWER

COMM POR

T

/

OUT

4 OUT

3

RELA

Y

2 RELA

Y

1

SENSOR INPUTS

C

NO

NC

NO

NC

C

C

NO

NC

NO

NC

C

N

L

+

-

DC ONLY

+

-

+

-

24 VDC

Regulated

Power Supply

V+

V-

No Connection:
Green
Blue
Brown

3-8900

3-2610-41

3-2610-41

Channel 1

Channel 2

White

Red

Red

Black

Black