For sipping:

1. Connect the sipper pump tubing to the outlet port on
the flow-through cuvette.

2. Secure the tubing using the clip on the righthand side
of the pump head.

3. Ease the tubing round the rollers by carefully rotating
them clockwise, by hand. Clamp the tubing into the clip
on the left hand side of the motor.

4. Once secured, ensure the tubing is routed into the two
retaining clips located on the base plate at the side of the
pump head.

5. Cut the tubing at the point where it fits comfortably
onto the left hand tube located on the inside of the front
bulk head.

6. Connect a suitable length of this tubing to the external
waste pipe.

7. Cut a small length of the sipper pump tube and push
this over one end of the capillary tube. Connect this to the
inlet port of the flow-through cuvette.

8. Route the tube into the two retaining clips located on
the base plate at the side of the pump head.

9. Fit the sipper probe and secure using the thumbscrew.
Feed the capillary tubing through the tube and up through
the sipper probe, allowing sufficient length for it to pass
into a suitable receptacle.

For pumping:

1. Cut two pieces of sipper pump tubing approximately
300mm in length. Take one length of tubing and fit this to
the pump head, as shown, securing the tubing using the
clip on the right hand side of the pump head.

2. Ease the tubing round the rollers carefully rotating them
clockwise, by hand. Clamp the tubing into the clip on the
left hand side of the motor.

3. Fit the other end onto the inlet port on the flow-through
cuvette.

4. Fit the second 300mm length of tubing to the outlet
port of the flow-through cuvette. Once secured, ensure
the tubing is routed into the two retaining clips located on
the base plate at the side of the pump head.

5. Fit the other end of the tubing onto the outlet port,
located on the inside of the front bulkhead.