OUTPUT SIGNALS
RS-232 / RS-485 Digital Output Signal
To use the RS-232 or RS-485 output signal, it is necessary to connect the RS-232

/ RS-485 Output Signal (Pin 5), the RS-232 / RS-485 Input Signal (Pin 3), and

Ground (Pin 8) to your computer serial port as shown on page 8. (See page 30

for details on accessing RS-232 / RS-485 output.)
Standard Voltage (0-5 Vdc) Output Signal
MC-Series fl ow controllers equipped with a 0-5 Vdc (opti onal 0-10 Vdc) will

have this output signal available on Pin 6. This output is generally available in

additi on to other opti onally ordered outputs. This voltage is usually in the range

of 0.010 Vdc for zero fl ow and 5.0 Vdc for full-scale fl ow. The output voltage is

linear over the enti re range. Ground for this signal is common on Pin 8.
Opti onal 0-10 Vdc Output Signal
If your controller was ordered with a 0-10 Vdc output signal, it will be available

on Pin 6. (See the Calibrati on Data Sheet that shipped with your controller to

determine which output signals were ordered.) This voltage is usually in the range

of 0.010 Vdc for zero fl ow and 10.0 Vdc for full-scale fl ow. The output voltage is

linear over the enti re range. Ground for this signal is common on Pin 8.
Opti onal Current (4-20 mA) Output Signal
If your controller was ordered with a 4-20 mA current output signal, it will

be available on Pin 1. (See the Calibrati on Data Sheet that shipped with your

controller to determine which output signals were ordered.) The current signal

is 4 mA at 0 fl ow and 20 mA at the controller’s full scale fl ow. The output

current is linear over the enti re range. Ground for this signal is common on Pin

8. (Current output units require 15-30Vdc power.)
Opti onal 2nd Analog Output Signal
You may specify an opti onal 2nd analog output on Pin 2 at ti me of order. (See

the Calibrati on Data Sheet that shipped with your controller to determine

which output signals were ordered.) This output may be a 0-5 Vdc, 0-10 Vdc, or

4-20 mA analog signal that can represent any measured parameter. With this

opti onal output, a controller could output the mass fl ow rate (0-5 Vdc on pin 6)

and the absolute pressure (0-5 Vdc on pin 2).

If your device is CSA/ATEX approved or equipped with the opti onal six pin

industrial connector, please contact Alicat.