CONTROL SETUP (conti nued)

LOOP VAR—The selecti on of what variable

to close the loop on is a feature unique to

Alicat mass fl ow controllers.
Pressing the LOOP VAR butt on on the

Control Setup screen will allow you to

change what variable is controlled.
Use the line-up and line-down butt ons

to move the arrow in front of the desired

opti on.
When the mass fl ow controller is supplied

with the control valve upstream of the

electronics porti on of the system, the unit

can be set to control on outlet pressure

(absolute pressures only) or volumetric

fl ow rate, instead of mass fl ow rate.
The change from mass to volume can

usually be accomplished without much, if

any, change in the P and D setti ngs.

When you change from controlling

fl ow to controlling pressure,

someti mes fairly radical changes

must be made to the P & D variables. See

page 20 – PID TUNING.

Contact Alicat if you are having diffi culti es

with this procedure.

ON AUTO / OFF AUTO—refers to the standard auto-tare or “auto-zero” feature.
The auto-tare feature automati cally tares (takes the detected signal as zero) the

unit when it receives a zero set-point for more than two seconds.
A zero set-point results in the closing of the valve and a known “no fl ow”

conditi on. This feature makes the device more accurate by periodically

removing any cumulati ve errors associated with drift .

It is recommended that the controller be left in the default auto-tare ON

mode unless your specifi c applicati on requires that it be turned off .

LN-DN

CANCEL

SET

LN-UP

>Mass Flow

Volumetric Flow

Pressure

LOOP

VAR

ON

AUTO

PID

MAIN

SETPT

SOURCE

SETPT

+0.0