Maintenance and Recalibration

General: MC-Series Flow Controllers require minimal maintenance. They

have no moving parts. The single most important thing that affects the life

and accuracy of these devices is the quality of the gas being measured. The

controller is designed to measure CLEAN, DRY, NON-CORROSIVE gases.
Moisture, oil and other contaminants can affect the laminar flow elements.

We recommend the use of in-line sintered filters to prevent large particulates

from entering the measurement head of the instrument. Suggested maximum

particulate sizes are as follows:

5 microns for units with FS flow ranges of 0-1 sccm or less.

20 microns for units with FS flow ranges between 0-2 sccm and 0-1 slpm.

50 microns for units with FS flow ranges of 0-1 slpm or more.

Recalibration: The recommended period for recalibration is once every year. A

label located on the back of the controller lists the most recent calibration date.

The controller should be returned to the factory for recalibration within one

year from the listed date. Before calling to schedule a recalibration, please note

the serial number on the back of the instrument. The Serial Number, Model

Number, and Date of Manufacture are also available on the Model Info display

(page 26).
Cleaning: MC-Series Flow Controllers require no periodic cleaning. If necessary,

the outside of the controller can be cleaned with a soft dry cloth. Avoid excess

moisture or solvents.