29 data recorder, 1 data recorder icon – Metrohm IC Net 2.3 User Manual

Page 274

Advertising
background image

6 Devices

IC Net 2.3

264

<Save as>

Opens the dialog for entering the path and file name.

Saves the last recorded spectrum in the selected

.cvs-file. Format: 2 lines of floating numbers (first line:

wavelength, second line: absorption or intensity).

<Copy>

Copies plot into clipboard.

<Close>

Closes

window.

If a right-hand mouse click is made in the spectrum window the wavelength

Current wavelength

and the corresponding absorption value

Current spec.

value

will be shown.

6.29 Data

recorder

6.29.1

Data recorder icon

The Data recorder icon is one of the components always pre-
sent in a new

SYSTEM

window. If the system is connected and

this icon is clicked with the right mouse button, the following
menu appears:

Open

Load the processing method linked to the sys-
tem and open an empty chromatogram window
(this window can also be opened by double-
clicking the icon).

Setup

Open data recorder setup window for selection
of processing method and data source (see be-
low).

Upload buffer

Upload of last chromatogram which was ac-
quired with the 762 IC Interface (

Upload buffer

doesn't work for all other interfaces). The chroma-
togram is immediately drawn on screen and
saved under a new name.

Unlink

Delete

the

Data recorder icon

from the system.

To record several single-channel chromatograms at the same
time, data recorder icons must be installed for each detector. To
record multi-channel chromatograms, only one data recorder
icon must be installed, but several data source channels have to
be linked to this data recorder.