10 Batch reprocessing

IC Net 2.3

358

10.3

Batch reprocessing queue editor

10.3.1 Open batch reprocessing queue editor window

REPROCESS /

<Edit sample table>

This menu item opens the batch reprocessing editor program for
editing the batch reprocessing table.

No

Row

number.

File name

Name of the chromatogram file (read-only).

Method

Name of the method file

*.mtw

used for re-

cording the chromatogram (read-only).

Ident

User defined identifier for chromatogram.
Will be placed into appropriate passport
field when starting the chromatogram.

Vial

Autosampler vial position to take sample
from. Will be placed into appropriate pass-
port field.

Volume

Injected volume of sample in

µ

L. Will be

placed into appropriate passport field.

Dilution

Sample dilution prior to injection. Will be
placed into appropriate passport field.

Amount

Sample amount. If this value is different for
the calibration run (c) and the sample run
(s), the component concentrations of the
sample are calculated as follows:

C

s

= C

c

Amount

s

/

Amount

c

Will be placed into appropriate passport field.

Internal standard amount

Concentration of the internal standard
component for relative concentration calcu-
lations. Will be placed into appropriate
passport field.

Calibration level

Calibration level (see section 7.5) for the
sample: Level

0

stays for normal analysis,

levels

1

and greater for calibration runs.

Correct filling of this column enables to cal-
culate calibration coefficients automatically.