5.3 830 IC Interface

IC Net 2.3

55

Program

On the

Program

subpage, program steps including time, program

instruction and parameter can be entered.

First column

Time at which program instruction is ap-
plied.
Entry range:

0.0 ... 999.9 min

If no time is entered, the program instruc-
tion is applied together with the last instruc-
tion with time entry.

Second column

Program instruction (see below).
In addition to these predefined instructions,
user-defined remote commands can be en-
tered if activated on the

Events configuration

tab.

Third column

Parameter for program instruction (see be-
low).

ENABLED

Enable program start (a disabled program
is not started).

<Add>

Add new program instruction.

<Delete>

Delete selected program instruction.

<Verify>

Test the time program (error messages are
displayed if program is wrong).

List of program instructions

The following program instructions can be added to the time pro-
gram on the

Program

subpage:

Instruction Parameter

entry Meaning

Events

0, 1, p, *

Set event output lines 1...7 to the

desired values. For entry of the first

value, enter

1

,

0

,

p

or

*

. For entry of

the other values, move the cursor in

front of the value to be changed

and enter

1

,

0

,

p

or

*

.

Program END,

RESET

The

END

flag can be used to end a

program, especially if the program

time should be longer than the

chromatogram duration. Additional

steps after this flag are not allowed.

The

RESET

flag is used to reset the

parameters to the system startup

values.