Defibtech DDU-100 Series User Manual

Page 22

Advertising
background image

14

DAC-510E-EN-BG

Within moments of insertion (if a non-discharged 9V ASI battery is installed) the DDU-100 will turn
on and run a battery pack insertion self-test. The unit will automatically shut off after the test is
run. Afterwards, the Active Status Indicator on the top corner of the DDU-100 AED will periodically
flash (if a nondischarged 9V ASI battery was previously installed in the battery pack). If the
indicator flashes green, the AED and battery pack are functioning properly, if the indicator flashes
red, there is a problem. Refer to the “Checking DDU-100 AED Status” section for more details on
the meaning of the indicator.

3.5 Connecting the Pads

The DDU-100 AED defibrillation/monitoring pads are supplied sealed in
a pouch with the connector and part of the cable exposed. This allows
the pads to be stored in a pre-connected state for rapid deployment
during an emergency.

Caution: DO NOT remove the defibrillation pads from the sealed
package until the pads are to be used. The packaging should be
opened only immediately prior to use, otherwise the pads may dry out
and become non-functional.

Note: The DDU-100 AED is designed to be stored with the pads
connector already installed. This simplifies the procedure for setting up
and operating the device in an emergency.

First, check to ensure that the pad package has not expired. Pads past

their expiration date should not be used and should be discarded.

Insert the connector end of the defibrillation pad cable into the pads connector port on the top-left
corner of the DDU-100 AED as shown. Insert pads connector firmly until it is fully seated in the unit.

The connected pad package can then be stored in the pad storage slot in the back of the DDU-100
AED. After connecting the pads connector to the unit, push the pad package, with the pictures
on the package facing out, rounded end first, into the pad holder compartment on the back of the
AED. When the pad package is fully inserted, press the pad cable into the groove in the back of
the unit to hold it in place and tuck any excess cable behind the pad package.

Caution: The pads are intended for one time use only and must be discarded after use or if the
package has been opened.

Advertising