4.9 Post Use Procedures

After the DDU-100 AED has been used on a patient, the unit should be cleaned following
procedures in the “Cleaning” section and prepared for the next use. The following steps should be
performed:

• Remove battery pack.
• Remove DDC if installed. Replace with a new DDC.
• Reinsert battery pack. Check that the Battery Pack Insertion Self-Test passes.
• Connect a new pad package (check to make sure the package is not expired).
• Hold ON/OFF button down for at least five seconds to initiate a manually initiated Self-Test.

Unit will report status of the Self-Test and shut off.

• Check to make sure that the Active Status Indicator is flashing green.

4.10 Operational Environment

The Defibtech AED is designed to operate in a wide range of environmental conditions. To ensure
the reliability and safety of the AED in a given environment, refer to the “Environmental” section
for a detailed list of approved environmental conditions.