8.1.6.2 Methods

AEDs were prospectively randomized according to defibrillation waveform on a daily basis in four
emergency medical services systems. First responders used either the 150-J biphasic AEDs or 200- to
360-J monophasic waveform AEDs on victims where defibrillation was indicated. A sequence of up to
three defibrillation shocks was delivered: 150J-150J-150J for the biphasic units and 200J-200J-360J for
the monophasic units. Defibrillation was defined as termination of VF for > 5 seconds, without regard to
hemodynamic factors.

8.1.6.3 Results

Of 338 patients with an out-of-hospital cardiac arrest, 115 had a cardiac etiology, presented with
ventricular fibrillation, and were shocked with one of the randomized AEDs. There were no statistical
differences between the monophasic and biphasic groups in terms of age, sex, weight, primary structural
heart diseases, cause or location of arrest, bystanders who witnessed the arrest, or type of responder. A
summary of the results is presented in table below.

Biphasic Patients

Number (%)

Monophasic Patients

Number (%)

P Value

Defibrillation Efficacy:

1 shock

< 2 shocks

< 3 shocks

52/54 (96%)

52/54 (96%)

53/54 (98%)

36/61 (59%)

39/61 (64%)

42/61 (69%)

< 0.0001

< 0.0001

< 0.0001

Patients defibrillated

54/54 (100%)

49/58 (84%)

0.003

ROSC

41/54 (76%)

33/61 (54%)

0.01

Survival to Hospital Admission

33/54 (61%)

31/61 (51%)

0.27

Survival to Hospital Discharge

15/54 (28%)

19/61 (31%)

0.69

8.1.6.4 Conclusion

More patients were defibrillated with an initial biphasic shock than monophasic shock and ultimately the
biphasic waveform defibrillated at higher rates than the monophasic waveform. A higher percentage of
patients achieved Return Of Spontaneous Circulation (“ROSC”) after biphasic shocks. Rates of survival to
hospital admission and discharge did not statistically differ between the two waveforms.