3 acoustic or infrasound sensors, 4 wind speed, 5 wind direction – Guralp Systems CD1.1 User Manual

Page 77

Advertising
background image

Operator's Guide

The CD1.1 calibration value for seismic sensors (including mass
position channels) should be in units of nm/count. To convert
from an acceleration gain to a dsiplaycement gain, it is
necessary to divide by (2πf)², or 4π²f². We also need to
introduce a scale factor of 1000, leading to:

C=1000

S

G

4π²f²

— CD1.1 calibration value for seismic sensor

with acceleration response, where C is the calibration value in
nm/count, S is the digitiser sensitivity in µV/count, G is the gain
of the instrument in V/m/s, and f is the frequency of the
calibration point in Hz.

12.3 Acoustic or infrasound sensors

For sensors which measure acoustic/infrasound responses, the
sensitivity (in units of V/Pa or VPa

-1

) should be known. The

CD1.1 calibration value is expected to be in Pa/count, so to
convert:

C=g

S⋅10

6

G

— CD1.1 calibration for acoustic sensor, where C

is the calibration value in Pa/count, g is the gain of an optional
auxiliary signal conditioner (use 1 if no conditioner is present),
S is the digitiser sensitivity in µV/count and G is the gain of the
instrument in V/Pa.

12.4 Wind speed

For a wind speed sensor with an output that is given as V/m/s
(or Vm

-1

s), use:

C=g

S⋅10

6

G

— CD1.1 calibration for acoustic sensor, where C

is the calibration value in ms

-1

/count, g is the gain of an

optional auxiliary signal conditioner (use 1 if no conditioner is
present), S is the digitiser sensitivity in µV/count and G is the
gain of the instrument in V/m/s.

12.5 Wind direction

For a wind direction sensor with an output that is given as V/°
(volts per degree), use:

June 2010

77

Advertising