Carrier axis failure – Integra CROMA24 User Manual

Page 81

Advertising
background image

Troubleshooting

81

TROUBLE-

SHOOTING

Croma24 Service Manual

Symptoms

Possible cause

Solution

Table 4-1. Troubleshooting Table (cont).

Carrier Axis
Failure

• dirty (or

perform Slide Shaft

lubricated) slide

Cleaning procedure

shaft

• perform Servo

replace servo motor

Motor Winding
Resistance check

• check servo

bad bearings -

motor for smooth

replace servo motor

movement

• obstruction in

remove obstruction

path of carrier
(may or may not
be visible)

• dirty encoder

perform Encoder

strip

Strip Cleaning
procedure

• damaged

replace encoder

encoder strip

strip

• bad encoder

replace encoder

sensor

sensor

• worn carrier

replace carrier

bushings

bushings

• loose trailing

reseat trailing cable

cable connections

connections at the
MPCB and the
carrier assembly