Carrier axis failure – Integra CROMA24 User Manual
Page 81
Advertising
Troubleshooting
81
TROUBLE-
SHOOTING
Croma24 Service Manual
Symptoms
Possible cause
Solution
Table 4-1. Troubleshooting Table (cont).
Carrier Axis
Failure
• dirty (or
perform Slide Shaft
lubricated) slide
Cleaning procedure
shaft
• perform Servo
replace servo motor
Motor Winding
Resistance check
• check servo
bad bearings -
motor for smooth
replace servo motor
movement
• obstruction in
remove obstruction
path of carrier
(may or may not
be visible)
• dirty encoder
perform Encoder
strip
Strip Cleaning
procedure
• damaged
replace encoder
encoder strip
strip
• bad encoder
replace encoder
sensor
sensor
• worn carrier
replace carrier
bushings
bushings
• loose trailing
reseat trailing cable
cable connections
connections at the
MPCB and the
carrier assembly