Counter n source signal, Routing a signal to counter n source, Counter n source signal -27 – National Instruments NI 6232 User Manual

Page 100: Routing a signal to counter n source -27

Advertising
background image

Chapter 7

Counters

© National Instruments Corporation

7-27

NI 6232/6233 User Manual

Counter n Source Signal

The selected edge of the Counter n Source signal increments and
decrements the counter value depending on the application the counter is
performing. Table 7-3 lists how this terminal is used in various
applications.

Routing a Signal to Counter n Source

Each counter has independent input selectors for the Counter n Source
signal. Any of the following signals can be routed to the Counter n Source
input.

80 MHz Timebase

20 MHz Timebase

100 kHz Timebase

RTSI <0..7>

Input PFI <0..5>

PXI_CLK10

PXI_STAR

In addition, Counter 1 TC or Counter 1 Gate can be routed to
Counter 0 Source. Counter 0 TC or Counter 0 Gate can be routed to
Counter 1 Source.

Some of these options may not be available in some driver software.

Table 7-3. Counter Applications and Counter n Source

Application

Purpose of Source Terminal

Pulse Generation

Counter Timebase

One Counter Time Measurements

Counter Timebase

Two Counter Time Measurements

Input Terminal

Non-Buffered Edge Counting

Input Terminal

Buffered Edge Counting

Input Terminal

Two-Edge Separation

Counter Timebase