Figure 4-14. reference trigger final buffer, Using a digital source, Ai pause trigger signal – National Instruments NI 6232 User Manual

Page 57: Ai pause trigger signal -30, Pause trigger signal, Section

Advertising
background image

Chapter 4

Analog Input

NI 6232/6233 User Manual

4-30

ni.com

Figure 4-14. Reference Trigger Final Buffer

Using a Digital Source

To use ai/ReferenceTrigger with a digital source, specify a source and an
edge. The source can be any of the following signals:

Input PFI <0..5>

RTSI <0..7>

PXI_STAR

The source also can be one of several internal signals on your DAQ device.
Refer to Device Routing in MAX in the NI-DAQmx Help or the
LabVIEW 8.x Help for more information.

You also can specify whether the measurement acquisition stops on the
rising edge or falling edge of ai/ReferenceTrigger.

Routing AI Reference Trigger Signal to an Output
Terminal

You can route ai/ReferenceTrigger out to any output PFI <6..9> or
RTSI <0..7> terminal.

AI Pause Trigger Signal

You can use the AI Pause Trigger (ai/PauseTrigger) signal to pause and
resume a measurement acquisition. The internal sample clock pauses while
the external trigger signal is active and resumes when the signal is inactive.
You can program the active level of the pause trigger to be high or low.

Reference Trigger

Pretrigger Samples

Complete Buffer

Posttrigger Samples