Figure 4-22. scanclk signal timing, Extstrobe* signal, Figure 4-23. extstrobe* signal timing – National Instruments NI 6014 User Manual
Page 56: Extstrobe* signal -30
Chapter 4
Connecting Signals
4-30
ni.com
Figure 4-22 shows the timing for SCANCLK.
Figure 4-22. SCANCLK Signal Timing
Note
When using NI-DAQ, SCANCLK polarity is low-to-high and cannot be changed
programmatically.
EXTSTROBE* Signal
EXTSTROBE* is an output-only signal that generates either a single pulse
or a sequence of eight pulses in the hardware-strobe mode. An external
device can use this signal to latch signals or to trigger events. In the
single-pulse mode, software controls the level of EXTSTROBE*. A 10
µs
and a 1.2
µs clock are available for generating a sequence of eight pulses in
the hardware-strobe mode.
Figure 4-23 shows the timing for the hardware-strobe mode
EXTSTROBE* signal.
Figure 4-23. EXTSTROBE* Signal Timing
Note
EXTSTROBE* cannot be enabled through NI-DAQ.
t
w
= 400 to 500 ns
t
d
= 50 to 100 ns
CONVERT*
SCANCLK
t
d
t
w
V
OH
V
OL
t
w
t
w
t
w
= 600 ns or 500
s