11.2 Technical Specifications – iRIS 350 / iRIS 350V

Dimensions:

117mm x 162mm x 67mm (4.60in x 6.37in x 2.64in)

(Excluding glands and mounting plate)

Mass:

1450g (3.19lb) with gel-cell battery fitted.

Power Supply:

Internal 12V @ 0.8A/Hr rechargeable sealed lead acid battery. Can also connect an
external regulated 12V dc power supply or rechargeable sealed lead-acid battery.

Charger Supply:

External 15-30Vdc supply. Supports a directly connected solar panel (no regulator).


Power Consumption:

Less than 3mA @ 12V in idle mode. Actual current consumption is dependent on
power management mode, wireless modem state and I/O configuration.

Comms Interfaces:

1x RS232 DB9, 38,400 bps, DTE configuration.
1x Integral multi-band GSM/GPRS modem or
1x Integral HSDPA (3G®) modem

Digital I/O:

4 x Digital Inputs/Outputs. Each channel configurable as either an input or output.

(Initial software fixes these as two inputs and two outputs – iRIS 320 emulation)

Digital Input Mode:
- 30Vdc maximum input, switch to 0Vdc to activate.
- Link selectable for either mechanical (<20Hz) or electronic (<5kHz) switching.

Digital Output Mode:
When set for output, the channel can be either:
- Switched 12V out (max 100mA) or
- Open-drain pull-down (max 100mA @ 30V)


Analog I/O:

4 x 16 bit uni-polar analog inputs. Range 0 –

5.000V. Input impedance 98kΩ.

Referenced to 0V common. Internal measurements available for monitoring are:

Battery Voltage

Supply Voltage

Internal Temperature (C or F)

Received Signal Strength (RSSI or RSCP)

Link selectable for either voltage (0-5V) or current (0-20mA) mode. Integral
current mode resistor value 100R.

SDI-12 Interface:

SDI-12 hardware interface that fully complies with the SDI-12 electrical standard.
Software support to SDI-12 standard 1.2.


Logging Memory:

Non-volatile 8MB flash storage of 1,085,476 time/date stamped data points. Circular
buffer mode (overwrites oldest data when memory full).
A typical site with 2 parameters logged every 15 minutes plus battery voltage logged
hourly will give 3.3 years of storage before data overwrite occurs.


Audio Memory:

iRIS 350V only. Non-volatile 8MB flash storage of PCM audio files. Total
cumulative playtime 12 minutes at 11.025kHz sampling rate.


Clock/Calendar:

Non-volatile with user replaceable lithium back-up battery module. Y2K compliant
with leap year recognition. Accurate to +/-20 secs month, trimmable via a
configuration register.


Mounting:

4 x M4 (3/16”) blind mounting holes in rear of case.
4 x 16mm (¾”) compression glands for power supply and I/O cabling access.


Environmental:

Storage Temperature: -20C - +85C. (-4F - +185F)

Operating Temperature: -10C - +70C. (14F - +158F)
Enclosure sealed to IP66 with gasket and glands.