Brooks, Model 5850i – Brooks Instrument 5850i User Manual

Page 43

Advertising
background image

4-15

Brooks

®

Model 5850i

Section 4 Maintenance

Installation and Operation Manual
X-TMF-5850i-MFC-eng
Part Number: 541B108AAG
December, 2008

4-6 Use of Orifice Sizing Nomograph

The Orifice and Restrictor Sizes for 5850i should be sized using the
"Brooks Thermal Mass Flowmeter Sizing Selection Program" Revision 8.6

or later. A copy can be requested through your local Brooks Sales
Representive or through the Brooks Customer Service Department.

The Orifice Sizing Nomograph, Table 4-4, is used to calculate the control
valve's orifice size when changing any or all of the following factors from
the original factory calibration.

gas
operating pressure (inlet and outlet)
flow range

The flow controller's orifice is factory-sized to a preselected gas, operating
pressure and flow range. Note that the orifice is marked with its size in
thousandths of an inch. When changing the aforementioned factors,
calculate the new orifice size by following the procedure and example
outlined below.

Example: Determine the orifice size for the following conditions:

Gas: Hydrogen
Flow Rate: 2000 sccm
Outlet Pressure: 30 psig
Inlet Pressure: 50 psig

1. Determine air equivalent flow rate (refer to Table 4-3).

where SG

AIR

= 1.00

Q

Air

=

Air equivalent flow rate (sccm)

Q

gas

=

Desired flow rate of the gas (sccm)

(Based on 0°C Standard temperature)

D

Air

=

Density of Air at 70°F

D

gas

=

Density of the gas (taken at customer temperature)

SG

gas

=

Specific gravity of the gas (taken at customer

temperature)

Refer to Table 4-3 for specific gravities.