Table A-2.
Accuracy Table of SCE Levels and Corresponding SPI Thresholds
References:
[1] Eftestol T, Sunde K, Steen PA. Effects of Interrupting Precordial Compressions on the Calculated Probability of
Defibrillation Success during Out-of-Hospital Cardiac Arrest. Circulation 2002; 105:2270-2273.
[2] Sota Y, Weil MH, Sun S, Tang W. Xie J, Noc M, Bisera J. Adverse effects of interrupting precordial compression
during cardiopulmonary resuscitation. Critical Care Medicine 1997; 25:733-736.
[3] Young C, Bisera J, Gehman S, Snyder D, Tang W, Weil MH. Amplitude spectrum area: measuring the probability of
successful defibrillation as applied to human data. Critical Care Medicine 2004; 32:S356-S358.
[4] Wik L. Rediscovering the importance of chest compressions to improve the outcome from cardiac arrest.
Resuscitation 2003; 58:267-269.
SCE Level
SPI Threshold
(mV-Hz)
Sensitivity
%
Specificity
%
HIGH
7.4
95
57
MEDIUM
8.3
90
65
LOW
10.8
80
80