ZOLL E Series Monitor Defibrillator Rev R User Manual

Page 124

Advertising
background image

E Series Operator’s Guide

A-26

9650-1210-01 Rev. R

Table A-2.

Accuracy Table of SCE Levels and Corresponding SPI Thresholds

References:

[1] Eftestol T, Sunde K, Steen PA. Effects of Interrupting Precordial Compressions on the Calculated Probability of
Defibrillation Success during Out-of-Hospital Cardiac Arrest. Circulation 2002; 105:2270-2273.

[2] Sota Y, Weil MH, Sun S, Tang W. Xie J, Noc M, Bisera J. Adverse effects of interrupting precordial compression
during cardiopulmonary resuscitation. Critical Care Medicine 1997; 25:733-736.

[3] Young C, Bisera J, Gehman S, Snyder D, Tang W, Weil MH. Amplitude spectrum area: measuring the probability of
successful defibrillation as applied to human data. Critical Care Medicine 2004; 32:S356-S358.

[4] Wik L. Rediscovering the importance of chest compressions to improve the outcome from cardiac arrest.
Resuscitation 2003; 58:267-269.

SCE Level

SPI Threshold

(mV-Hz)

Sensitivity

%

Specificity

%

HIGH

7.4

95

57

MEDIUM

8.3

90

65

LOW

10.8

80

80