Defibrillator – ZOLL E Series Monitor Defibrillator Rev R User Manual

Page 95

Advertising
background image

Troubleshooting Guidelines

9650-1210-01 Rev. R

13-5

Defibrillator

Symptom

Recommended Action

Excessive artifact when using paddles as ECG source.

Ensure PADDLES is selected.

Firmly press paddles against patient skin.

Use gel on paddles.

Clean paddle surface.

Check and clean between adult and pediatric shoe.

Check cable for damage.

Use MFE electrodes for ECG analysis.

Defibrillator will not charge (energy level does not
increment on display).

Check SHOCK button(s) on paddles or front panel are
not stuck on.

Replace battery with a fully charged battery pack.

Charge time to 200 J exceeds 10 seconds.

Typical in a low battery condition (up to 20 seconds)

Change battery pack.

Plug device into AC power.

Install fully charged battery pack.

Energy does not discharge when you press SHOCK
button(s).

60 seconds have elapsed in Manual mode since initial
charge ready. Energy was internally discharged.

15 seconds have elapsed in automatic mode since
initial charge ready. Energy was internally discharged.

Device is in SYNC mode and no QRS complex is
detected.

Energy internally discharged because energy selection
was changed during charge or after the device was
ready.

Unit not completely charged when you pressed SHOCK
button(s). Wait for DEFIB XXXJ READY message and
ready tone.

Press and hold SHOCK button(s) until energy is
delivered to the patient.

Device is connected to the AutoPulse Plus and no
relaxation period of the compression cycle has been
detected yet.

Unit is unable to SHOCK when in SYNC mode.

Ensure SYNC XXXJ SEL message is displayed on
monitor.

Check for SYNC marker above R wave. If not present,
change ECG size, lead selection, or electrode
placement.

Press and hold SHOCK button(s) until energy is
delivered to the patient.

Alter ECG electrode placement.

Make sure ECG signals are displayed.

Advertising