Restarting the device – ZOLL E Series Monitor Defibrillator Rev R User Manual
Page 22
E Series Operator’s Guide
1-14
9650-1210-01 Rev. R
Restarting the Device
Certain events require the E Series products to be
restarted after they shut off or become inoperative.
One example is when the battery runs down and the unit
shuts off. In this case, perform these steps in this order:
1. Turn the selector switch to the OFF position
2. Remove the battery.
3. Insert a new battery.
4. Turn the selector switch to the desired operating
mode to resume operation.
This sequence is necessary to restart the device, and
can also be used to clear some X FAULT XX messages,
when immediate use of the device is required.
Note that some settings (for example, alarm settings,
lead selection, ECG size) may need to be restored from
their default values when operation is resumed.