E Series Operator’s Guide

13-6

9650-1210-01 Rev. R

No energy delivery to patient is apparent.

Under certain circumstances, some patients do not
twitch when energy is delivered.

Perform defibrillator self test as described in “General
Maintenance” on page 11-1.

Check for CHECK PADS and POOR PAD CONTACT
messages alternating on the monitor.

If Multi-Function electrodes are used, ensure proper
placement and contact.

Unit displays PADDLE FAULT message when
connected to the AutoPulse Plus.

Discharge energy internally by changing the energy
selection.

Disconnect the Multi-Function Cable and PADS from the
AutoPulse Plus, and connect the Multi-Function Cable
directly to the PADS.

Unit displays the CHECK PADS prompt.

Verify proper Multi-Function Cable/MFE Pad connection
by disconnecting and reconnecting the Multi-Function
cable and MFE Pads.

Ensure proper contact of Multi-Function Pads and that
the patient does not have excessive hair beneath the
electrodes.

If message persists, disconnect Multi-Function cable
from MFE Pads and plug cable into test connector.
CHECK PADS prompt should change to DEFIB PAD
SHORT
(Manual mode only).

If test fails, try using paddles to defibrillate.

Unit displays the USE PADS prompt.

The ECG analysis function operates only when MFE
pads are attached to the patient. You must either:

Disconnect paddle and connect MFE Pads for use in
Semiautomatic defibrillation.

Activate Manual mode to use paddles.

Unit displays the NOISY ECG or the
RETRY ANALYSIS prompt.

Check for proper application and adhesion of Multi-
Function electrodes.

Check to make sure that nobody is touching the patient
and that the patient is motionless.

Unit displays the ECG TOO LARGE, or the RETRY
ANALYSIS
prompt.

Press ANALYZE button again to begin analysis.

No TEST OK message appears when performing a
defibrillator self-test.

Check to make sure unit is set to 30 Joules.

If testing with Multi-Function Cable, make sure that
cable is firmly inserted into test connector.

If testing with paddles, make sure to press the paddles
firmly against the sides of the unit while discharging.

Unit displays the DEFIB MAINT. REQUIRED message.

Contact ZOLL Technical Service Department.

Unit displays a DEFIB FAULT XX message.

If problem persists, contact ZOLL Technical Service
Department.

Symptom

Recommended Action