Yokogawa Touch Screen GP20 User Manual

Page 359

Advertising
background image

5-5

IM 04L51B01-01EN

Maintenance and T

roubleshooting

1

2

3

4

5

App

Note

Do not perform any other operations (especially turning the GX/GP off) during calibration.

The GX/GP may malfunction.

Using a Password

You can use a password to allow only permitted users to perform calibration.

1

Tap the

MENU key, Browse tab, Initialize Calibration, and menu A/D calibration.

The A/D calibration call-up screen appears.

Enable the password (On).

Set the password.

2

Tap

On/Off under Password settings to select On.

3

Tap

Password, and set the password.

The password is displayed using asterisks.

From the next time, enter the password to show the A/D calibration call-up screen.

Enter the password.

4

Tap

Exit to exit from the Initialize Calibration screen.

Operation complete

Note

• Default password: default

• Initialization does not initialize the A/D calibration password.

If you forget the password, you will not be able to perform A/D calibration. Make sure you do not

forget the password.