B.91 [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPlitude] COMMAND

VOLT:TRIG

Syntax:

Short Form: [SOUR:]VOLT[:LEV]:TRIG[:AMP] <exp_value>
Long Form: [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPlitude] <exp_value>
<exp_value> = digits with decimal point and Exponent, e.g., 2.71E1 for 27.1

Description: Programs voltage value to be transferred to output by *TRG command. Actual output voltage will

depend on load conditions. If the value exceeds the maximum for the model being programmed, error
message -222,”Data out of range” is posted in output queue. If value exceeds VOLT:LIM:HIGH value,
a value corresponding to the voltage limit will be programmed.

B.92 [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPlitude]? QUERY

VOLT:TRIG?

Syntax:

Short Form: [SOUR:]VOLT[:LEV]:TRIG[:AMP]?
Long Form: [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPlitude]?
Return Value: <exp_value> = digits with decimal point and Exponent, e.g., 2.71E1 for 27.1

Description: Returns value representing voltage value to be programmed by *TRG command established by

VOLT:TRIG command).

B.93 STATus:OPERation:CONDition? QUERY

STAT:OPER:COND?

Syntax:

Short Form: STAT:OPER:COND?

Long Form: STATus:OPERation:CONDition?

Return Value: <int_value> 0 to 1313 (1 + 32 + 256 + 1024).

Description: Returns the value of the Operation Condition Register (see Table B-3). The Operation Condition

Register contains unlatched real-time information about the operating conditions of the power supply.
Bit set to 1 = function enabled (active, true); bit reset to 0 = function disabled (inactive, false). (See
example, Figure B-4.)

B.94 STATus:OPERation:ENABle COMMAND

STAT:OPER:ENAB

Syntax:

Short Form:

STAT:OPER:ENAB <int_value> 0 to 1313 (1 + 32 + 256 + 1024)

Long Form:

STATus:OPERation:ENABle <int_value> 0 to 1313 (1 + 32 + 256 + 1024)

Description: Sets Operation Enable Register. The Operation Enable Register is a mask for enabling specific bits

in the Operation Event Register which will cause the operation summary bit (bit 7) of the Status Byte
register to be set Bit set to 1 = function enabled (active, true); bit reset to 0 = function disabled (inac-
tive, false). The operation summary bit is the logical OR of all the enabled bits in the Operation Event
register. (See example, Figure B-4.)

B.95 STATus:OPERation:ENABle? QUERY

STAT:OPER:ENAB?

Syntax:

Short Form: STAT:OPER:ENAB?

Long Form: STATus:OPERation:ENABle?

Return Value: <int_value>

Description: Reads Operation Enable Register (see Table B-3). Returns value of Operation Enable Register bits.

Bit set to 1 = function enabled (active, true); bit reset to 0 = function disabled (inactive, false). (See
example, Figure B-4.)

B.96 STATus:OPERation[:EVENt] QUERY

STAT:OPER?

Syntax:

Short Form: STAT:OPER[:EVEN]?

Long Form: STATus:OPERation[:EVENt]?

Return Value: <int_value>

Description: Indicates changes in conditions monitored by Operational Event Register (see Table B-3).

Returns the value of the Operation Event register. The Operation Event register is a read-only register
which holds (latches) all events that occur. Reading the Operation Event register clears it. (See exam-
ple, Figure B-4.)

TABLE B-3. OPERATION CONDITION REGISTER, OPERATION ENABLE REGISTER,

AND OPERATION EVENT REGISTER BITS

CONDITION

NU

CC

NU

CV

NU

NU

NU

NU

BIT

15-11

10

9

8

7 - 6

5

4 - 1

0

VALUE

32,768 -

2048

1024

512

256

128 - 64

32

16 -2

1

CC -

POWER SUPPLY IN CONSTANT CURRENT MODE

CV -

POWER SUPPLY IN CONSTANT VOLTAGE MODE

NU

NOT USED