Sect. 7.6 Part C Special applications, functional checks, service, and order numbers

IFC 210 E

05/2002

72


Check of setpoint display

1)

Switch on power supply, allow at least 15 minutes’ warm-up time.

2)

Set switch D (front panel of GS 8A) to “0” position.

3)

Adjust zero to 0 or 4 mA with the 10-turn potentiometer P (front panel of GS 8A),

depending on setting in Fct. 1.05, deviation < ±10 µA.

4)

Calculate position of switch Y and displayed setpoints ”I“ and ”f“:

Q

100%

×

K

4.1)

X =

GK

×

DN

2

Q

100%

full-scale range (100%) in volumetric unit V per unit time t

GK

primary constant, see instrument nameplate

DN

meter size DN im mm, not inches, see instrument nameplate

t

time in seconds (Sec), minutes (min) or hours (hr)

V volumetric

unit

K constant, according to following table

:

t

Sec min hr

V

Liter

25 464

424.4

7.074

m

3

25 464 800

424 413

7 074

US-Gallonen

96 396

1 607

26.78

4.2) Determine position of switch Y: Use table (front panel GS 8A) to determine value Y,

which comes closest to factor X and meets condition Y

X.

4.3) Calculate setpoint reading ”I“ for current output:

I = I

0%

+ (I

100%

- I

0%

) in mA

I

0%

current (0/4mA) at 0% flowrate

I

100%

current (20mA) at 100% flowrate

4.4) Calculate setpoint reading ”f“ for pulse output:

f =

×

P

100%

in Hz

P

100%

pulses per second (Hz)

at 100% flowrate

5)

Set switch D (front panel GS 8A) to position “+“ or “-“ (forward/ reverse flow).

6)

Set switch Y (front panel GS 8A) to the value determined by the method described above.

7)

Check setpoint readings I and f, see points 4.3 and 4.4 above.

8)

Deviation < 1.5% of setpoint. If greater, replace signal converter, see Sect. 8.3.

9)

Test of linearity: set lower Y values, readings will drop in proportion

to the calculated Y values.

10)

Switch off power supply after completing the test.

11)

Disconnect the GS 8A.

12)

Reassemble in reverse order, see points e) – b) under “electrical connection”,

see illustration in Sect. 8.1.

13)

The system is ready for operation after the power supply has been switched on.

Example: see overleaf