Yamato Scientific HC200 Cool Block User Manual
Page 35

33
How to release from the state after the power restoration
If a blackout, which includes a instantaneous blackout, occurs during the quick auto-stop operation
and the program operation with the resume function activated, in case of the quick auto-stop
operation, HC200 starts the quick auto-stop operation from the first and in case of the program
operation, HC200 runs the fixed temperature operation at the preserved temperature.
When you select the operation mode again in this state, you need HC200 released from the state after
the power restoration by operating below. Note that you can not release HC200 from the present
state if you only turn off and on the power switch. Because, to turn off the power switch makes the
controller recognized that it is in a blackout.
Indication after Step/Step Procedure
Explanation
The display after the power restoration
when the resume function has been set to on
Ex. ;The temperature
in the chamber>
Ex.;The temperature
in the chamber
•
When the power restored, the display shown on the
left appears when the resume function has been set to
on.
There are three ways to release from the above state.
Indication after Step/Step Procedure
Explanation
1. Use the abort function
Press the MODE key
and push either the
▲ key or the ▼ key
•
Press the MODE key and push either the ▲ key or the
▼ key until
appears on the main display.
•
Then press the ENTER key.
2. Per form the fixed temperature operation
The preceding set
temperature in the fixed
temperature operation
Ex. ;The temperature
in the chamber>
Press MENU key
and change to the
fixed temperature
operation and then
press ENTER key.
•
Change to the fixed temperature operation mode by
pushing the MENU key and press the ENTER key.
When once you change to the program operation,
HC200 can release from the state after the power
restoration.
3. Per form the program operation
Ex. ;Select the program #2
Select the program
number you need by
pushing either the
▲ key or the ▼ key
and press the
ENTER key
•
Select the program number you need by pushing
either the ▲ key or the ▼ key and press the ENTER
key.
•
The program operation that you have selected starts
from the first.