How to release from the state after the power restoration

If a blackout, which includes a instantaneous blackout, occurs during the quick auto-stop operation
and the program operation with the resume function activated, in case of the quick auto-stop
operation, HC200 starts the quick auto-stop operation from the first and in case of the program
operation, HC200 runs the fixed temperature operation at the preserved temperature.
When you select the operation mode again in this state, you need HC200 released from the state after
the power restoration by operating below. Note that you can not release HC200 from the present
state if you only turn off and on the power switch. Because, to turn off the power switch makes the
controller recognized that it is in a blackout.

Indication after Step/Step Procedure

Explanation

The display after the power restoration

when the resume function has been set to on

Ex. ;The temperature
in the chamber>

Ex.;The temperature
in the chamber

When the power restored, the display shown on the

left appears when the resume function has been set to
on.

There are three ways to release from the above state.

Indication after Step/Step Procedure

Explanation

1. Use the abort function

Press the MODE key

and push either the
▲ key or the ▼ key

Press the MODE key and push either the ▲ key or the

▼ key until

appears on the main display.

Then press the ENTER key.

2. Per form the fixed temperature operation

The preceding set

temperature in the fixed

temperature operation

Ex. ;The temperature
in the chamber>

Press MENU key

and change to the

fixed temperature

operation and then

press ENTER key.

Change to the fixed temperature operation mode by

pushing the MENU key and press the ENTER key.
When once you change to the program operation,
HC200 can release from the state after the power
restoration.

3. Per form the program operation

Ex. ;Select the program #2

Select the program

number you need by

pushing either the

▲ key or the ▼ key

and press the

ENTER key

Select the program number you need by pushing

either the ▲ key or the ▼ key and press the ENTER
key.

The program operation that you have selected starts

from the first.