3.5 Configuring instrument components

■■■■■■■■■■■■■■■■■■■■■■

22

■■■■■■■■

864 Robotic Balance Sample Processor

2 Enter the configuration data

Enter the following data:

– Swing offset –8.0°
– Maximum swing range 105°
– Swing radius 110 mm
– Rotation offset 0.0°
– Swing direction +

3.5.2

Configuring the towers

Various lift positions must be defined before methods are created for the
864 Robotic Balance Sample Processor.

On Tower 1 settings are necessary for:

Lift position for picking up a pipetting tip

Lift position for aspirating and discharging sample liquid.

Robotic arm and lift positions for deflecting the pipetting tips.

On Tower 2 settings are necessary for:

Lift position for titrating

Lift position for rinsing the titration head on the rack
or

Robotic arm and lift position for rinsing the titration head in the wash-
ing station

Tower 1

Configure the positions for tower 1:

1 Open the tower configuration

Click the Configuration icon.

Double-click the instrument name 864_1 in the Devices win-
dow.

Click the tab Tower 1.

2 Enter the configuration data

Enter the following settings:

– Max. stroke path 215 mm
– Swing position 136 mm
– External position 1, angle 127.9°
– External position 1, work position 136 mm
– External position 2, angle 138.2°
– External position 2, work position 129 mm