15 mounting the drip pan, Figure 29, Mounting the tubing to the drip pan – Metrohm 864 Robotic Balance Sample Processor User Manual

Page 52

Advertising
background image

3.15 Mounting the drip pan

■■■■■■■■■■■■■■■■■■■■■■

44

■■■■■■■■

864 Robotic Balance Sample Processor

7

Swing Head cable
Connection cable of the 786 Swing Head
Connector on tower 1 (Swing Head)

8

Sensor connection
Electrode cable (plug F2 m)

Sensor (on tower 2)

— Titrando (Ind.)

9

USB connection
6.2151.000 controller cable (USB A-MiniDIN
8p)

864 (USB)

— Titrando (Controller)

10 Dosino connection (pipetting drive P)

Dosino connection cable
Connector on 864 (MSB 1)

11 Dosino connection (titrant T)

Dosino connection cable
Connector on Titrando (MSB 1)

12 Dosino connection (solvent S)

Dosino connection cable
Connection to the 864 Robotic Balance
Sample Processor (MSB 2)

3.15

Mounting the drip pan

Serious damage to the instrument or a danger to the user can occur if
chemicals or liquid samples are spilled. The use of the drip pan
6.2711.080
is recommended in order to avoid such incidents.

Figure 29

Mounting the tubing to the drip pan

First, fasten the tubing enclosed to the drainage nipple on the drip pan
and then guide the free end of the tubing into a waste container.

Install the drip pan as follows:

Advertising