3 Installation

864 Robotic Balance Sample Processor

■■■■■■■■

39

2 Connect the cable

Connect the mains cable and the RS 232 cable (with RJ-45 plug)
to the rear side of the balance.

Fasten the grounding cable to the rear side of the balance.

Pull the cable through under the housing of the Sample Processor
from the right and guide it to the rear.

Positioning the balance

The balance provided comes equipped with a clamping ring and a weigh-
ing plunger. The weighing plunger may not be attached at the time of the
installation. Proceed as follows:

1

2

3

Figure 25

Installing the balance

1 Position the clamping ring

Slightly loosen the clamping ring.

Guide the clamping ring over the weighing column.

2 Position the balance

Slide the balance from the right-hand side underneath the hous-
ing of the 864 Robotic Balance Sample Processor.
The clamping ring must fit into the recess of the housing in such a
way that the position of the balance can be readily fastened into
place with the clamping ring.
Do not tighten the clamping ring too firmly. The position of the
balance needs to be finely aligned later on.