Figure 14, Mounting the pipetting tip adapter – Metrohm 864 Robotic Balance Sample Processor User Manual
Page 35
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
864 Robotic Balance Sample Processor
■■■■■■■■
27
6.1808.250
6.1808.000
6.1805.130
1
2
3
4
Figure 14
Mounting the pipetting tip adapter
1 Position the adapter
■
Loosen the screw and the washer of the adapter and guide the
adapter into the robotic arm head from below.
2 Fix the adapter
■
Screw the adapter tightly with the screw and the washer. If nee-
ded, tighten carefully with a wrench.
3 Attach the tubing adapter
■
Screw the 6.1808.000 tubing adapter (with 2x M6 inner thread,
supplied with the 6.1808.250 adapter) tightly onto the adapter.
4 Connect the tubing
■
Manually tighten the 6.1805.130 PTFE tubing (120 cm) to the
tubing adapter and place it into the guide chain of the tower.
Mounting the robotic arm with the titration head
After initialization, the drive disc of the Swing Head is positioned as
though the robotic arm were located in the outermost position.
Mount the 6.1462.260 robotic arm to tower 2 as follows: