3.4 Connecting external instruments
Metrohm 838 Advanced Sample Processor, Installation
39
The Metrohm remote cables have a designation at the ends of the
cable, indicating for what instrument the relevant connector is sched-
uled and at what connector position it must be connected.
Example:
732/1 Remote
3.4.2 Serial connections (RS232)
The connection options at the serial RS232 interface are diverse. Be-
sides all Metrohm instruments featuring the Metrohm remote control
language, it is possible to connect a personal computer. Any third-party
instruments featuring a serial RS232 interface may also be connected.
Correct data transfer requires correct setting of the transmission pa-
rameters. They must correspond to the settings of the interface on the
connected instrument.
3.4.3 Remote
interface
Connected peripheral instruments, such as the 819 Advanced IC De-
tector etc., can be controlled via the remote interface (25-pin socket).
14 lines (output 0–13) are available for output of signals.
8 lines (input 0–7) are available for reception of signals.
Only the designated Metrohm remote cables should be used to con-
nect Metrohm instruments.