4.2 Care and maintenance

64

Metrohm 838 Advanced Sample Processor, Service, maintenance, faults


Please use only the specified

6.2744.000 PVDF pressure screws

for connection to the ultrafiltration cell. If you use the 6.2744.010

PEEK pressure screws, this may result in stress cracking on the ultrafil-

tration cell!

4 Closing the ultrafiltration cell

• Fit the lower section 45 onto the upper section 43 so that the

two parts fit perfectly.

• Using the 6.2621.070 Allen key, fully screw in and firmly

tighten the 5 screws 46.

5 Fitting the ultrafiltration cell

• After screwing the ultrafiltration cell together, insert it into the

holder 26 as shown in Figure 17 so that the heads of the

screws 46 are located in the holes provided in the holder.

• Fit the three capillary connections which were detached pre-

viously back on the inlet and outlet openings provided on the

ultrafiltration cell using the PVDF pressure screws 22.

6 Rinsing the ultrafiltration cell

• Each time the ultrafiltration membrane is replaced, any air still

in the filtration cell and in the lines must be removed. In order

to do this, rinse all lines with ultra-pure water from one of the

special beakers for instance:

• Place a 300 ml PE bottle with ultra-pure water as the rinsing

solution onto a special beaker position on the sample rack.

• Use the IC Net software to control the 838 Advanced Sample

Processor directly. Choose the corresponding system and

open the Manual control dialog box. You can run the indi-

vidual functions of the Sample Processor under Autosam-

pler. Please refer to the IC NET software Online Help

documentation for details of operation.

• Move the beaker with the rinsing solution above the lift with

MOVE.

• You can use the LIFT function to lower the needle into rinsing

position.

• Switch on the peristaltic pump with PUMP, and all connected

lines are then rinsed and any air they contain is removed.

• After an adequate time (e.g. 5 min), switch the pump back off

again with PUMP and move the needle back to home posi-

tion.