3.5 Connection to Metrohm IC systems

Metrohm 838 Advanced Sample Processor, Installation

43

001 Valve

fill

002 Move

spec1

003 Lift

special

004 Dosino MSB1 rate 5
005 Dosino MSB1

3.5

006 Move

sample

007 Dosino MSB1 rate -0.1
008 Lift

work

009 Dosino MSB1 rate 5
010 Dosino MSB1

-3

011 Move

spec1

012 Lift

special

013 Dosino MSB1

compen.

014 Wait

5

015 Valve

inject

016 Wait

1

017 Dosino MSB1 rate 2
018 Dosino MSB1

Concentr.vol.

019 Wait

5

020 Valve

fill

021 Dosino MSB1 rate 5
022 Dosino MSB1

3.5

023 Valve

inject

024 Wait

1

025 Dosino MSB1 rate 2
026 Dosino MSB1

Dilution vol.

027 Stirrer MSB2

15

028 Valve

fill

029 Ctrl

Fill A1

030 Pump speed

4

031 Pump

65

032 Ctrl

Zero 1

033 Ctrl

Inject A1

034 Pump speed

10

035 Pump

30

036 Stirrer MSB2

on

037 Valve

inject

038 Dosino MSB1 rate 20
039 Dosino MSB1

5

040 Pump

on

041 Valve

fill

042 Wait

60

043 Pump

off

044 Valve

inject

045 Dosino MSB1 rate 20
046 Dosino MSB1

5

047 Valve

fill

048 Pump speed

10

049 Pump

on

050 Wait

60

051 Stirrer MSB2

off

052 Wait

10

053 Pump

off

054 Dosino MSB1

fill

− Switch over injection valve on 838 to "Fill"

− Move needle to rinsing beaker *

− Position lift with needle to special position *

− Set dosing rate

− Eject 3.5 ml of ultra-pure water (prepare pipet.)

− Move needle to sample

− Form air bubble

− Move lift with needle to working position *

− Set dosing rate

− Aspirate 3 ml of sample **

− Move needle to rinsing beaker

− Move lift with needle to special position

− Compensate for mech. play of Dosino coupling

− Waiting time

− Move injection valve on 838 to "Inject" position

− Waiting time

− Set dosing rate

− Sample volume is dosed

− Waiting time

− Switch over injection valve on 838 to "Fill"

− Set dosing rate

− Eject rest of sample into rinsing beaker

− Set injection valve on 838 to "Inject" position

− Waiting time

− Set dosing rate

− Dilution volume (ultra-pure water) is dosed

− Set stirring speed

− Switch over injection valve on 838 to "Fill"

− Switch over injection valve on 820 to "Fill"

− Set delivery rate

− Transfer diluted sample to sample loop

− Trigger autozero on 819/1 IC Detector

− Switch over injection valve on 820 to "Inject"

− Set delivery rate

− Mixing vessel is emptied

− Switch on stirrer

− Set injection valve on 838 to "Inject" position

− Set dosing rate

− Rinse mixing vessel with ultra-pure water

− Switch on pump

− Switch over injection valve on 838 to "Fill"

− Empty mixing vessel again

− Switch off pump

− Set injection valve on 838 to "Inject" position

− Set dosing rate

− Rinse mixing vessel again

− Switch over injection valve on 838 to "Fill"

− Set delivery rate

− Switch on pump

− Empty mixing vessel again

− Switch off stirrer

− Waiting time

− Switch off pump

− Fill Dosino for next dilution

* must be predefined in «IC Net»

** max. volume must be taken into account. The sample must

not reach the Dosino cylinder.