Measurement setup examples, Eccentricity/vibration measurement setting methods, Peak-to-peak function 2. timing input – KEYENCE LK-G5000 Series User Manual

Page 10

Advertising
background image

10

5

Measurement setup examples

Eccentricity/Vibration Measurement Setting Methods

Setting methods when performing eccentricity/vibration measurements are as follows.

This section provides application setup examples.

Preparation and Overview

When measuring eccentricity/vibration for flat or
cylindrical workpieces, the “Peak-to-peak function” of
the LK-G Series is used.

The peak-to-peak function automatically calculates the difference between MAX and MIN value for a custom sampling
cycle. The custom sampling cycle can be controlled using the timing input or by RS-232C commands.

The target is rotated, time input is turned ON, and sampling is started. When the timing input is turned ON once
again, sampling ends and the maximum variation in the measurement value is displayed, making it possible to obtain
comparator output.

1. Peak-to-peak function

2. Timing input

Required Components

No. of units

LK-G5000 Series controller

1

LK-G5000 Series sensor head

1

LK-G5000 Series head cable

1

LK-Navigator 2 configuration software

1

24 V power source

1

Controller

Display of

vibration

amount

(V)

ON

OFF

(t)

Start of measurement

Timing signal

End of measurement

MAX value

Output value

MIN value

Comparator output

Sampling time

Timing signal

0.031

Comparator output