Bcd output switching, Reset period and refresh cycle – KEYENCE LS-7600 User Manual

Page 161

Advertising
background image

7-13

Chapter 7 I/O Terminals

7

BCD Output Switching

ON

OFF

BCD SELECT input

ON

OFF

BCD DATA output

ON

OFF

BCD INDICATE output

OUT1 Measured value

ON

OFF

OUT1 STROBE output

OUT2 STROBE output

ON

OFF

ON

OFF

OUT2 Measured value

Refresh cycle

Refresh cycle

Max2.1ms

OUT1

OUT1

OUT1

OUT1

OUT1

OUT2

OUT2

Max2.1ms

0

±

0.1ms

• BCD output can be set to OUT1 or OUT2 according to BCD SELECT input.
• OUT1 and OUT2 are refreshed asynchronously according to the user settings for the

averaging number and measuring mode. OUT1 or OUT2 can be selected according
to BCD SELECT input.

• It is possible to use BCD INDICATE output to check if the present BCD output is

OUT1 or OUT2.

Reset Period and Refresh Cycle

The following table shows the relationship of the averaging number, reset time, and the
periods of refresh cycle.
If the measuring mode is set to sample hold 2 or self-timing hold, change the item
Reset Time with Sampling Period.

Number of

averaging times

Reset time [ms]

Refresh cycle

[ms]

1

0.42

0.42

2

0.83

0.83

4

1.67

0.83

8

3.33

0.83

16

6.67

0.83

32

13.33

0.83

64

26.67

0.83

128

53.33

0.83

256

106.67

1.67

512

213.33

3.33

1024

426.67

6.67

2048

853.33

13.33

4096

1706.67

26.67