Temperature characteristics – KEYENCE LS-7600 User Manual
Page 199
9-5
Chapter 9 Specifications
9
LS-7070M
150
56
40
65
Receiver
Transmitter
Unit : mm
20
30
100
±
50
±
6
μ
m
±
3
μ
m
Temperature Characteristics
The following table shows typical examples of the temperature characteristics of the
LS-7600 Series.
LS-7010M
Temperature
Drift
0˚C (32˚F)
0.25
50˚C (122˚F)
-0.15
10˚C (50˚F)
0.10
20˚C (68˚F)
0.00
40˚C (104˚F)
-0.10
30˚C (86˚F)
-0.05
Unit:
μ
m
❈
The above values are checked with a 1 round bar located in the center of the meas-
uring range at a reference temperature of 20˚C.
LS-7030M
Temperature
Drift
0˚C (32˚F)
–0.5
50˚C (122˚F)
0.2
10˚C (50˚F)
–0.1
20˚C (68˚F)
0.0
40˚C (104˚F)
0.1
30˚C (86˚F)
0.1
Unit:
μ
m
❈
The above values are checked with a ø10 round bar located in the center of the
measuring range at a reference temperature of 20˚C.
LS-7070M
Temperature
Drift
0˚C (32˚F)
+1.8
50˚C (122˚F)
-3.0
10˚C (50˚F)
+0.8
20˚C (68˚F)
0.0
40˚C (104˚F)
-1.8
30˚C (86˚F)
-0.8
Unit:
μ
m
❈
The above values are checked with a ø20 round bar located in the center of the
measuring range at a reference temperature of 20˚C.