KEYENCE LS-7600 User Manual
Page 188
8-26
Chapter 8 RS-232C
8
Communication using an External Synchronous Trigger
The following section provides information on how to obtain output from the RS-232C
interface in synchronization with a timing input into the input terminal.
This operation does not require ASCII commands. Therefore, use this function if the KV-10/
80 Series (or the KV mode) is used only to retrieve measuCommunication using an Exter-
nal Synchronous Triggered values or a data logger with no Communication using an
External Synchronous Trigge/command transmission function is connected to the LS-7600
Series.
Note:
• No commands are available for measured value output or setting changes.
The printer cannot be connected while the LS-7600 Series is under external
synchronous control.
• While the RS-232C interface is transmitting measurements, do not trigger
the next measured value, otherwise the next measured value will not be
output.
Connections
■
Environment Settings
Make the following environment settings. Refer to RS-232C on page 6-7 for environ-
ment settings in detail.
Item
BAUDRATE
9600bps
STOPBIT
D-LENGTH
D-MODE
D-SEND
1
PARITY
EVEN
8
PLC
S1/S2
See notes below.
Set value
Remarks
The settings specified in this table
will be automatically selected
when the D-MODE item is set to
PLC.
❈
When the D-MODE item is set to PLC, the STX+ETX characters will be selected for
delimiter insertion.
❈
When the D-SEND item is set to S1 or S2, measured value output under external
synchronous control will be possible.
■
Output
Output All Measured Values
With the D-SEND item set to S1, all measured values will be output.
Output NG (No Good) Measured Values
If the D-SEND item is set to S2, NG measured values will be output.