8-26

Chapter 8 RS-232C

8

Communication using an External Synchronous Trigger

The following section provides information on how to obtain output from the RS-232C
interface in synchronization with a timing input into the input terminal.
This operation does not require ASCII commands. Therefore, use this function if the KV-10/
80 Series (or the KV mode) is used only to retrieve measuCommunication using an Exter-
nal Synchronous Triggered values or a data logger with no Communication using an
External Synchronous Trigge/command transmission function is connected to the LS-7600
Series.

Note:

• No commands are available for measured value output or setting changes.

The printer cannot be connected while the LS-7600 Series is under external
synchronous control.

• While the RS-232C interface is transmitting measurements, do not trigger

the next measured value, otherwise the next measured value will not be
output.

Connections

Environment Settings

Make the following environment settings. Refer to RS-232C on page 6-7 for environ-
ment settings in detail.

Item

BAUDRATE

9600bps

STOPBIT

D-LENGTH

D-MODE

D-SEND

1

PARITY

EVEN

8

PLC

S1/S2

See notes below.

Set value

Remarks

The settings specified in this table
will be automatically selected
when the D-MODE item is set to
PLC.

When the D-MODE item is set to PLC, the STX+ETX characters will be selected for
delimiter insertion.

When the D-SEND item is set to S1 or S2, measured value output under external
synchronous control will be possible.

Output

Output All Measured Values
With the D-SEND item set to S1, all measured values will be output.

Output NG (No Good) Measured Values
If the D-SEND item is set to S2, NG measured values will be output.