Fluke Biomedical ESA615 User Manual
Page 49
Advertising
![background image](/files/785315/content/doc049.png)
Electrical Safety Analyzer
How to Do Electrical Safety Tests
35
The outlet conditions below apply when you do this test:
•
Normal
Polarity
•
Normal Polarity, Open Neutral
•
Reversed
Polarity
•
Reversed Polarity, Open Neutral
IEC60601-1 specifies that the applied parts must be
connected for this measurement. Push or to ground
and unground all applied parts connection posts.
Note
A ground connection on an applied part is shown
as a box around the applied part in the display.