Alternative Equipment Leakage Test
Note
The alternative equipment leakage test is
available when the EN62353 standard is
selected.
The Alternative Equipment Leakage test, applies the
voltage source between short-circuited equipment outlet
mains live, neutral, and equipment outlet earth, the
exposed conductive surface on the housing, and all
applied parts short-circuited together. The test disconnects
equipment from mains. The current which flows over the
insulation of the DUT is measured.
This test is not applicable for equipment with internal
electrical power source. The switches in mains part must
be closed for this measurement.
To do an alternative equipment leakage test:
1. Push
.
2. Push
the Alternative Equipment softkey.
3. Push
to apply the voltage and read the current
in the display.
Figure shows the electrical connections between the
Product and the DUT for the Alternative Equipment
Leakage Test.
The outlet conditions below apply when you do this test:
•
Closed
Earth
•
Open
Earth
Note
If there are more than five applied parts to
connect to the Product, see the How to Use the
1-to-10 Adapter section in this manual.
Alternative Applied Part Leakage Test
Note
The Alternative applied part leakage test is
available when the EN62353 standard is
selected.
The Alternative Applied Part Leakage test applies the test
voltage between short-circuited applied parts of a single
function and the short-circuited equipment outlet mains
live, neutral, earth, and exposed conductive surface on the
housing. This test should only be done for equipment with
F-Type applied parts. For equipment with multiple applied
parts, test each group of applied parts of a single function
in turn with all others float during the test. All applied parts
can be connected to the applied parts jacks and the lead
selection will float those not selected.