Detailed specifications – Fluke Biomedical ESA615 User Manual
Page 93
Electrical Safety Analyzer
Detailed Specifications
79
Detailed Specifications
Test Standard Selections ........................................ ANSI/AAMI ES-1, IEC62353, IEC60601-1, and AN/NZS 3551
Voltage
Ranges (Mains voltage) ...................................... 90.0 V to 132.0 V ac rms
180.0 V to 264.0 V ac rms
Range (Point-to-point voltage)
5000 m ............................................................ 0.0 V to ≤150 V ac rms
2000 m ............................................................ 0.0 V to ≤300.0 V ac rms
Accuracy..............................................................
±
(2 % of reading + 0.2 V)
Earth Resistance
Modes.................................................................. 2-Wire
Test Current ........................................................ >200 mA ac
Range .................................................................. 0.000
Ω
to 2.000
Ω
Accuracy..............................................................
±
(2 % of reading + 0.015
Ω
)
Equipment Current
Range .................................................................. 0.0 A to 20.0 A ac rms
Accuracy..............................................................
±
(5 % of reading + (2 counts or 0.2 A, whichever is greater))
Duty cycle ............................................................ 15 A to 20 A, 5 min on/5 min off
10 A to 15 A, 7 min on/3 min off
0 A to 10 A, continuous
Leakage Current
Modes* ................................................................ AC+DC (True-rms)
AC only
DC only
* For tests that do not use MAP voltage, AC+DC, AC ONLY, and DC ONLY modes are
available for all leakages. MAP voltages are available only in True-rms (shown as
AC+DC)