Electrical Safety Analyzer

How to Do Point-To-Point Measurements

59

How to Do Point-To-Point Measurements

The Product can make voltage, resistance, and low
current measurements through its Point-to-Point function.
To access the Point-to-Point function menu shown in
Figure , push . Softkeys F1 through F3 are used to
set the measurement function.

gtv128.jpg

Figure 34. Point-To-Point Function Menu

Measure Voltage

To make a voltage measurement:
1. Push

the

Voltage softkey from the Point-To-Point

menu.

2. Put test leads in the RED (V/

Ω

/A ) and BLACK jacks.

3. Put the probe tips across the unknown voltage and

read the measurement in the display.

The Product measures a maximum of 300 volts ac.

Measure Resistance

To make a resistance measurement:
1. Push

the

Resistance softkey from the Point-To-Point

menu.

2. Put test leads in the RED (V/

Ω

/A ) and BLACK jacks.

3. Null lead resistance by shorting the leads together

and Push the Zero Leads softkey.

4. Put the probes across the unknown resistance and

read the measurement in the display.

The Product measures resistances to a maximum of
2.0

Ω

.