Electrical Safety Analyzer
How to Do Point-To-Point Measurements
59
How to Do Point-To-Point Measurements
The Product can make voltage, resistance, and low
current measurements through its Point-to-Point function.
To access the Point-to-Point function menu shown in
Figure , push . Softkeys F1 through F3 are used to
set the measurement function.
gtv128.jpg
Figure 34. Point-To-Point Function Menu
Measure Voltage
To make a voltage measurement:
1. Push
the
Voltage softkey from the Point-To-Point
menu.
2. Put test leads in the RED (V/
Ω
/A ) and BLACK jacks.
3. Put the probe tips across the unknown voltage and
read the measurement in the display.
The Product measures a maximum of 300 volts ac.
Measure Resistance
To make a resistance measurement:
1. Push
the
Resistance softkey from the Point-To-Point
menu.
2. Put test leads in the RED (V/
Ω
/A ) and BLACK jacks.
3. Null lead resistance by shorting the leads together
and Push the Zero Leads softkey.
4. Put the probes across the unknown resistance and
read the measurement in the display.
The Product measures resistances to a maximum of
2.0
Ω
.