7.2
HART/analog single loop wiring
Note
For HART communications:
•
600 Ω maximum loop resistance
•
250 Ω minimum loop resistance
HART/analog single loop wiring
Figure 7-2:
A
B
A.
820 Ω maximum loop resistance
B.
HART-compatible host or controller
I/O wiring for Model 1700 and Model 2700 transmitters with analog outputs