C A L I B R A T I O N A N D T R O U B L E S H O O T I N G
Recommended Equipment

90

LDX-3232

C H A P T E R

6

Finally, the LDX-3232 Current Source should be allowed to warm up for at least
one hour before calibration.

Recommended Equipment

Recommended test equipment for calibrating the LDX-3232 Current Source is
listed in Table 6.1. Equipment other than that shown in the table may be used if
the specifications meet or exceed those listed.

Table 6.1 Recommended Test Equipment

It will be necessary to connect various loads and circuits to the outputs of each
current source for the calibration procedure. A schematic is shown in Figure 6.1
for the photodiode calibration circuit with the required components listed in Table
6.2. Also, the devices required for other calibration loads are listed in Table 6.2.

Table 6.2 Required Calibration Components

Description

Manufacturer/Model

Specification

DMM

HP 3457A

DC Amps (@ 1.0 A): ±0.02%
Resistance (@ 10

Ω): ±0.02%

0.1 µA or 0.1 mV resolution

Laser Current Source Calibration

Current Source Calibration, all models

10

Ω, 20 W resistor, low TCR

Voltage Measurement Calibration

15

Ω, 5 W resistor, low TCR

PD Monitor Calibration (See Schematic in Figure 6.1)

PD Calibration Circuit

R1

R2

R4

R3

U1

49

Ω resistor, 1%, 1/4 W

100

Ω resistor, 1%, 1/4 W

1.0 M

Ω resistor, 1%, 1/4 W

5

Ω, 1%, 10W

TIL 117 opto isolator