C A L I B R A T I O N A N D T R O U B L E S H O O T I N G

Local Calibration of the LDX-3232 Current Source

04_06

LDX-3232

91

C H A P T E R

6

Figure 6.1 I

PD

Calibration Circuit

Local Calibration of the LDX-3232 Current Source

There are three calibration adjustments required for the LASER current source of
the LDX-3232 Current Source. They are calibration of the constant current source
for both bandwidths and ranges, calibration of the laser voltage measurement,
and calibration of the constant light power (I

PD

) feedback circuits.

The LDX-3232 Current Source implements a two-point calibration for the Laser
current source. Two currents (approximately 80% and 20% of FS) are applied to
a test load, and the resulting actual currents are fed back (by the user) to the
Current Source. The Laser Current Source calibration program uses the two sets
of data to calculate new calibration constants. These calibration constants are
used to set the actual current output of the current source.

Current Source Calibration

The following procedure is for calibrating each range of the current source.
Calibration must be performed on both bandwidth modes. A total of four
calibrations will be performed on the current source. Start with the low range, low
bandwidth mode. Configure the Laser Current Source in the following state:

Function

State

MODE

I (constant current low bandwidth)

RANGE

low (lowest per instrument model)

CURRENT LIMIT

90% of FS (full scale of range)

SETPOINT

80% of FS

U 1 TIL117

1

2

3

4

5

6

A

V

R 3

R2

100

9-Pin D-Sub

LD C athode (5)

LD Anode (9)

PD C athode + (6)

PD Anode - (7)

Voltm eter

Am m eter

R1

49

Ipd C urrent

R4
1 M

CA LIB R ATIO N C IR C U IT