4 technical data, Technical data, Instrument components and specifications – LEICA ASP300 User Manual

Page 15

Advertising
background image

15

Leica ASP300

3.4 Technical data

Nominal voltage:

Two factory-preset voltage settings
(not user-adjustable):
100 to 120 V + 1 0% or
230 to 240 V + 1 0%

Nominal frequency:

50 to 60 Hz

Main fuses:

2 fast-blowing melting fuses, 20 x 5 mm, UL-listed

Order no.

100 to 120 V; F 10 A 250 VAC

69411002

230 to 240 V; F 5 A 250 VAC

69411000

Nominal power:

1,000 VA

Dimensions (W x D x H), in mm:

595 x 680 x 1,325

Dry weight, unpacked:

approx. 140 kg

Weight, packed:

220 kg

Operating temperature range:

15 °C to 35 °C

Relative humidity:

10% to 80% - non-condensing

IEC 1010 classification:

Protective Class 1
Pollution Degree 2
Overvoltage Installation Category II:

800 V impulse (120 V systems)

1500 V impulse (240 V systems)

Maximum operating altitude:

maximum 2,500 m

Local / remote alarm relays:

30 V DC, maximum 2 A

2 connections:
both single voltage-free, change-over contacts
(can be operated as both normally-open and
normally-closed connections)

Paraffin wax baths

Number of baths:

3

Volume (liters):

4.3 l per bath

Melt time:

approx. 10 hrs

Temperature:

40 to 65 °C

Temperature accuracy:

+ 1 °C

3.

Instrument components and specifications