Operation – LEICA ASP300 User Manual

Page 33

Advertising
background image

33

Leica ASP300

Fig. 16

Adding molten paraffin

Molten wax may have a maximum temperature of 70 ° C.

If filling with molten wax, do not fill above the MAX level (see

Fig. 15

).

After a wax bath has been filled, it must be marked as ‘full’ in the
REAGENT STATUS screen.

Software control

If the wax baths are filled with molten wax, it is possible to override
the Leica ASP300 melting time calculation.

For that purpose, in supervisor mode, access the
SERVICE FUNCTIONS screen (

Fig. 16

).

Press TIME UNTIL WAX MELTED.

Enter the required delay time and press OK
to confirm.

If the instrument attempts a paraffin step with the paraffin only partly molten, severe instru-
ment malfunctions may occur.

5.

Operation

Advertising