Instrument components and specifications – LEICA ASP300 User Manual

Page 16

Advertising
background image

16

Instruction manual V 1.3 – 12/2003

Retort

Capacity:

max. 300 cassettes

Reagent volume (in liters):

4.3

Temperature (paraffin):

40 to 65 °C

Temperature (processing reagents):

ambient or 35 to 55

o

C

Temperature (cleaning reagents):

50 to 65 °C

Temperature accuracy:

+ 1 K

Filling time:

approx. 90 sec

Drain time :

approx. 80, 120, 140 sec (adjustable)

Impregnation vacuum:

-70 kPa (g)

Impregnation pressure:

35 kPa (g)

Fill vacuum:

-70 kPa (g)

Drain pressure:

35 kPa (g)

General

Reagent bottles

10

Cleaning solution bottles:

3

Maximum bottle volume:

4.3 l

Printer:

Optional accessory

Pretest check:

ON/OFF

Fluid level sensor:

ON/OFF

Recirculation (pump in/out):

ON/OFF

(a) Time before 1st cycle:

12 min

(b) Time between cycles:

20 min

System setup
Password status:

Supervisor / User

Type of password:

alphanumeric, freely selectable

Reagent Management Status:

ON/OFF

Software access control (‘Lock Mode’):

ON/OFF

3.

Instrument components and specifications

Advertising