Operation – LEICA ASP300 User Manual

Page 35

Advertising
background image

35

Leica ASP300

Program Options

Selection of the options on how programs are carried out.

STATION ALLOCATION:

By age or Sequential

By age

-

RMS is activated, reagents will automatically be
used in order of increasing cleanliness.

Sequential -

reagents will be used in sequential order of stations.

WAX BATH ORDER:

Auto or 1.; 2. ; 3.

Wax bath order can be activated only when STATION ALLOCATION
is set to Sequential.

PROMPT FOR NUM BLOCKS :

Enabled or Disabled

Enabled-

When starting a program, the actual number of
blocks must be entered.
This is recommended with the RMS being enabled.

Disabled -

Number of blocks can be entered optionally (not
mandatory).

Date / Time

Be sure to verify that the date and time entered do actually correspond
to the local time/date.

Press the buttons to access the entry windows where time/date can be
corrected, if necessary.

Instrument

Press WAX BATH SET TEMP (°C) to access the wax bath temperature
entry window. Select the highest temperature compatible with the type
of paraffin used as paraffin temperature will drop slightly once the paraf-
fin is in the retort.

5.

Operation

Advertising